InPro 6800 G Amperometric Oxygen Sensor
Published on : Tuesday 30-11--0001
METTLER TOLEDO’s line of amperometric gas-phase oxygen sensors are well suited for a wide measurement range in a broad choice of applications. These in-line sensors provide excellent performance with minimum maintenance and do not require any gas sampling and pre-conditioning systems.
Features and Benefits
In-situ measurement
- Measurements are fast and are performed directly in the process stream.
Accurate
- Insensitive to moisture and dust, the membrane – covered sensing element is permeable to oxygen, not liquids.
Reduced service time
- Thanks to the easy-to-replace membrane body and quick-disconnect interior body.
Long sensor lifetime
- Durable and rugged sensor design for increased resistance to harsh environments.
Installation in hazardous zones
ATEX certificate: SEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69oC/T81oC/T109oC/ T161oC Da/Db
IECEx Certificate of Conformity: IECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69oC/T81oC/T109oC/ T161oC Da/Db
FM approval: IS/ I, II, III /1 /ABCDEFG /T6
Specifications - InPro 6800 G Amperometric Oxygen Sensor
Accuracy |
≤ ± [1% +0.1 Vol-%] |
Measuring pressure resistance |
0.2 to 9 bar (2.9 to 130 psi absolute) |
Operating range |
0.1 Vol-% O? to 100 Vol-% O? |
90% response time at 25°C (77°F) |
≤ 20 s (N2®15 Vol-% O?) |
Measuring principle |
Polarographic Clark electrode |
Sensor body |
316L stainless steel or C22 (titanium on request) |
Membrane material |
PTFE/Silicone (reinforced with steel mesh) |
O-ring material |
Silicone or Kalrez® |
Mechanical pressure resistance |
Maximum 12 bar (174 psi absolute) |
Certificates |
SEV 14 ATEX 0169 X |
IECEx Certificate of Conformity |
IECEx SEV 14.0026X |
FM approval |
IS/ I, II, III /1 /ABCDEFG /T6 |
It is used in Oil And Gas, Chemical Industries.
Mettler-Toledo India Private Limited, Tel:022 4291 0256, Email:[email protected]