Flue Gas Analyzer System
Published on : Tuesday 30-11--0001
Continuous monitoring of up to 5 components:
NOx, SO2, CO, CO2, and O2
Ideal for flue gas monitoring in waste incinerators, boilers, and furnaces
- Space-saving design
- A variety of functions: auto calibration, O2-corrected output, averaging, alarm, etc.
- Japanese type approval
Specifications
Principle |
NOx, SO2, CO, CO2: Non dispersive infrared sensor |
|
---|---|---|
Components |
1 component: NOx, SO2, CO, CO2 |
|
Range |
NOx: 0–50 ppm…5000 ppm |
|
Repeatability |
±0.5% FS |
|
Zero drift |
≤ ±1.0% FS per week |
|
Span drift |
≤ ±2.0% FS per week |
|
Linearity |
±1.0% FS |
|
Response time |
≤ 120 s (for 90% response, from the analyzer inlet) |
|
Output signal |
• 4–20 mA DC (linear) |
|
Contact input |
Auto-calibration start, average reset, range switch, output hold, pump ON/OFF, etc. |
|
Functions |
Auto calibration, O2 correction, averaging, CO peak count and alarm, data recording |
|
Operating environment |
-5°C to +40°C or -10°C to +40°C, RH 90% or lower |
|
Power supply voltage |
100, 110, 115, 200, 230 V AC, 50/60 Hz |
|
Dimensions |
800 (H) × 1710 (W) × 615 (D) mm, for indoor use |
|
Gas conditions |
Temperature |
Standard: ≤ 800°C |
Dust |
≤ 100 mg/Nm3 |
|
Pressure |
-3 to +3 kPa |
|
Component |
SO2: ≤ 500 ppm |
It is used in chemical,Oil And Gas Industry.
Fuji Electric India Pvt. Ltd
Phone: +91-22-4010 4870
Email: [email protected]