Teledyne Gas Calibrators & ZAG- T753U
Published on : Tuesday 30-11--0001
The Model T753U is a portable, microprocessor based precision Calibrator for Ozone gas instruments and sensors. Using a sophisticated Ozone generation system, combined with a sensitive photometer and pneumatic design, the T753U produces stable and reliable concentrations of Ozone in a convenient and rugged package based on our popular T750/U Portable Multigas Calibrator platform. The T753U is housed in a high-density, molded travel case which includes an integral handle and wheels for durability and ease of use in the field for virtually any application. An internal zero air option extends the T753U's capabilities to also provide a clean and dry source of zero air, eliminating the need for a separate source of zero air. The internal zero air source module contains a software driven electronic inlet valve manifold allowing the calibrator to automatically switch between internal and external zero air sources depending on the application needs.?
The T753U Portable Ozone Calibrator's performance is characterized by our own Model T265 Chemiluminescence O3 Analyzer; a US EPA Federal Reference Method (FRM) for Ozone. With an interference-free Ozone measurement and best available lower detection limit (LDL) of < 0.3 ppb, the T753U performance is proven by the most sensitive Ozone gas analyzer technology available.
All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, and flexible I/0. All instrument set up, control and diagnostic information is available through the front panel, or via RS232, Ethernet, or the optional USB com port, either locally or by remote connection
The Model T753U comes with NumaView ™ Software. NumaView™ Remote PC software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView ™ software.?
Features
- Transfer standard for calibration of ozone analyzers
- Stable and accurate ozone output to 2 ppb
- Built-in ozone generator and photometer
- Large, vivid, and durable color graphics display with touch screen interface
- Ethernet, RS-232, and (optional) USB com ports
- Front panel USB ports for peripheral devices and firmware upgrades
- 12 independent timers for sequences
- Nested sequences (up to 5 levels)
- Flow adjustable from 2 to 5 lpm (with internal zero air source)
- Flow adjustable from 2 to 15 lpm (with external zero air source)
- Inlets for external ozone reference sources
- Internal Zero Air Source Module (optional)
- Two-year warranty?
Linearity |
± 1% of full scale |
Response Time |
< 240 seconds to 95% |
Stability (7 days) |
<1 % |
Flow Rate (with internal zero air source) |
2 to 5 LPM adjustable |
Flow Rate (with external zero air source) |
2 to 15 LPM adjustable |
Maximum Output |
2.5 ppm LPM |
Minimum Output |
10 ppb LPM |
Minimum Ozone Concentration |
2 ppb |
Response Time |
<240 seconds to 98% |
Range |
0 ppb to 10 ppm |
Precisi?on |
1.0 ppb |
Linearity |
±1% of full scale |
Rise/Fall Time |
<20 Seconds to 95% (photometer response) <180 seconds to 95% (system response)
|
Zero Drift |
<1.0 ppb/7 days |
Span Drift |
< 1%/24 hours
|
Lag time |
<10 seconds |
Flow Rate |
800 cc/min ±10% |
Power Requirements |
100V-120V, 220V-240V, 50/60 Hz |
Analog Ouput Ranges (Test Channel) |
10V, 5V, 1V, 0.1V (selectable) |
Included I/O |
1 x Ethernet: 10/100 Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
12 x digital control outputs
12 x digital control inputs
8 x digital status outputs
|
Optional I/O |
1 x USB com port
1 x RS485
Multidrop RS232
|
Operating Temperature Range |
5-40 ºC? |
Dimensions (HxWxD) |
9” x 17” x 21” (229 x 432 x 533 mm) |
We?ight Specifications are subject to change without notice.
All specifications are based on constant conditions.?
|
41.5 lbs (18.8 kg with internal zero air option)? |
It is used in Ambient Air Monitoring and Road Side Air Monitoring, and all chemical company.
Teledyne Technologies Incorporated.
San Diego, California,United States
Phone: (858) 657-9800
Email: [email protected]